General Specifications |
Dimensions: |
5.55" x 5.25" x 1.65" (14.1
cm x 13.40 cm x 4.15 cm) |
Weight: |
18.5 oz (574 g) |
Power Supply: |
Internal Battery P/N QSE 601 NiCad Battery
Pack; External Specifi ed Power Supply: P/N L60071 Medical Grade Power Supply. |
Channels: |
(4) for INF, NMS, and TENS (2) Active (red) & (1)
Dispersive (black) for High Voltage |
Electrodes/Leadwires: |
(4) Lead Wires - Style L00013MC (2) GV0014R
(red single wires) and (1) GV0014B (Black single wire) |
Tolerance: |
+/- 1% |
|
Interferential
Specifications |
Waveform: |
Continuous Sinusoidal |
CH1/CH3 (Carrier) Frequency:
|
4000 Hz |
CH2/CH4
Frequency: |
4001-4150
Hz adjustable |
Beat
Frequency: |
1-150 Hz
- CH1/CH2 are an Interferential Pair
- CH1: Fixed Frequency (4000 Hz)
- CH2: Adjustable 4001-4150 Hz
- CH3/CH4 are an Interferential Pair
- CH3: Fixed Frequency (4000 Hz)
- CH4: Adjustable 4001-4150 Hz
|
Premodulated
INF (Using one Channel): |
A mixed
Interferential Beat Frequency will be produced only out of CH 1 and/or CH 3. |
Modes: |
Constant
Stimulation
Alternating Stimulation
Cycled Stimulation |
Constant: |
Beat Frequency
is adjustable from 1-150 Hz |
Sweep
1: |
Beat Frequency
Modulates from 1-10 Hz and back over 12 seconds (6 second increase, 6 second
decrease) |
Sweep
2: |
Beat Frequency
Modulates from 80-150 Hz and back over 16 seconds (8 second increase, 8 second
decrease) |
Sweep
3: |
Beat Frequency
Modulates from 1-150 Hz and back over 20 seconds (10 second increase, 10 second
decrease. |
Intensity: |
0-78 milliamps
peak to peak |
Output
Voltage: |
0-39 Volts
peak to peak |
|
Neuromuscular Specifications |
Waveform: |
Symmetrical, biphasic, square Asymmetrical,
biphasic, square |
Pulse
Rate: |
Adjustable 1-120 Hz peak to peak |
Pulse Width: |
Adjustable 50-400 µS |
On Ramp: |
0-10 seconds |
On Time: |
0-99 seconds |
Off Ramp: |
0-10 seconds |
Off Time: |
0-99 seconds |
Modes: |
- Constant: Use the Constant Mode or a Pulse Rate greater than
60 Hz. for the relaxation of muscle spasms only. Use of continuous stimulation
and/or rates above the normal physiological range (e.g., greater than 60 Hz)
could lead to rapid onset of muscle fatigue, making the device less effective
in producing repeated forceful muscle contractions.
- Cycled
- Reciprocating
|
Intensity: |
0 - 120 mA from peak to peak (adj)Mo |
Voltage: |
0- 60 volts from peak to peak (adj.) |
|
TENS
Specifications |
Waveform: |
Asymmetrical Biphasic Square |
Pulse Rate: |
Adjustable 1-120 Hz |
Pulse Width: |
Adjustable 10-250 µS |
Modes: |
Constant
Pulse Rate (Hz) is adjustable from 1-120 Hz and Pulse Width (µS) is adjustable
from 10-250 µS.
Pulse Width Modulation:
Decrease/Increase of set width 50% over 5 second cycle.
Pulse Rate Modulation:
Decrease/Increase of set width 50% over 5 second cycle.
Pulse Width & Pulse Rate Modulation:
Pulse Rate and Pulse Width alternately decrease and increase 50% of their set
values over 5 seconds cycle.
Strength Duration 1:
Increase of set Pulse Width 40%, decrease of set Pulse Rate 45% and decrease
of Amplitude 10% over a 3 second period. Values return to original settings over
the next 3 second period.
Strength Duration 2:
Increase of a set Pulse Width 60% decrease of set Pulse Rate 90% and decrease
of set Amplitude 13% over a 6 second period. Values return to original setting
over the next 6 seconds.
Cycled Burst (B1):
2.5 seconds on, 2.5 seconds off. Adjust able Pulse Rate 1-200 Hz. Adjustable
Pulse Width from 10-250 μs.
Cycled Burst II (B2):
8 pulses per 0.25 seconds, off for 0.25 seconds. Adjustable Pulse Width from
10-250 μs. |
Intensity: |
0-100 mA peak to peak |
Voltage: |
0-50 Volts peak to peak |
|
High-Volt Specifications |
Waveform: |
Pulsed galvanic twin peak monophasic delivered in pairs |
Channels: |
2 active outputs
1 dispersive return |
Output Voltage: |
0 - 350 volts adjustable |
Output Current: |
0 - 700 mA peak, Adjustable |
Pulse Width: |
Inter pulse intervals of 100 (μS) ,fixed. |
Pulse Rate: |
1 - 120 Hz, adjustable |
Modes: |
Constant Stimulation
Alternating Stimulation
Cycled Stimulation |
Polarity Switch: |
Positive/Negative polarity of active electrodes
(A1 & A2) |
Ramp On: |
0-10 seconds |
On Time: |
0-99 seconds |
Ramp Off: |
0-10 seconds |
Off Time: |
0-99 seconds |