BIOSTIM® M7

BIOSTIM® M7
TENS UNIT
BioMedical Life Systems
www.bmls.com


Overview

BioMedical Life Systems Inc., Digital, BioStim® M7 TENS Device has seven programmable functions; Continuous, Width Modulation, Rate Modulation, Rate and Width Modulation, SD1, SD2 and Cycled Burst. The device has a Patient Lock System that, when activated prevents the patient from changing any of the set parameters. When the Patient Lock Systems is turned off, a Patient Compliance Meter is displayed showing how long the device was used. A timer allows the patient to use the device for a specified period of time. The two-channel device incorporates touch proof design and is run by two AA Batteries.

General Specifications

Size:
3.25” x 2.25” x 1 5/8” (82 mm x 70mm x 45mm)
Weight:
8 oz (266 grams)
Channels:
Dual
Waveform:
Asymmetrical rectangular biphasic (TENS)
Pulse Rate (adj.):
1- 200 Hz
Pulse Width (adj.):
10 - 250 µs
Stimulation Modes:
 
Continuous:
Constant Stimulation
Rate Modulation:
50% decrease/increase of set value over a 5 second cycle
Width Modulation:
50% decrease/increase of set value over a 5 second cycle
Rate/Width Modulation:
50% decrease in set value over a five second period. As rate increases, width decreases.
SD1:
Increase of set pulse width 40%, decrease of set pulse rate 45% and decrease of amplitude 10% over a 3 second period. Values return to original settings over the next 3 seconds.
SD2:
Increase of set pulse width 60%, decrease of set pulse rate 90% and decrease of amplitude 13% over a 6 second period. Values return to original settings over the next 6 seconds.
Cycled Burst:
2.5 seconds on. 2.5 seconds off. Adjustable Rate and Width.
Output Voltage:
49 Volts peak to peak
Intensity (adj.):
0-98 mA
Power Supply:
2 AA Batteries
 
Data was recorded across a 500 ohm resistance load.