We have electrotherapy down to a science
TRANSCUTANEOUS ELECTRICAL NERVE STIMULATOR
System 2000

Print

Overview

BioMedical Life Systems Inc. Systems 2000® is a Two channel analog TENS device with Five modes of operation: Continuous, Cycled Burst, 50% Modulation of set Width, 50% Modulation of set Rate and Modulation of set Width and Rate conversely.  Pulse Rate adjustable 2 to 200 Hz and Pulse Width is adjustable 10-250 us.  The unit is powered by one 9-volt battery.

Next
System 2000

Print

General Specifications

Size: 3.9" x 2.75" x 1" (9.90 cm x 6.98 cm x 2.54 cm)
Weight: 4.6 oz (132 grams)
Channels: Dual
Power Source: 9V battery
Waveform: Asymmetrical rectangular biphasic pulse
Output Current: Constant
Intensity (adj.): 0 - 100 mA peak
Output Voltage: 0 - 50V peak to peak
MODES  
Conventional: Pulse rate (adj.) 2 to 200 Hz
Pulse width (adj.) 10 to 250 µs
Pulse Rate Modulation: 50% decrease of set value
Pulse Width Modulation: 50% decrease of set value
Pulse Rate and Pulse - Width Modulation: As rate decreases, width increases
As rate increases, width decreases
over a five-second interval
Cycle Burst: Adjustable bursts cycled at 2.5 second intervals. Rate dial sets number of pulses within each burst
Number of Electrodes: 2 pair
Number of Lead Wires: 2 pair
Tolerances: +/- 10%
Output parameters are across a 500 ohm resistance.
Phone: 800-726-8367 (USA) | 760-727-5600 (Outside USA) | 760-727-4220 (Fax)
Email: information@bmls.com (general) | sales@bmls.com (Order Information)
BioMedical Life Systems, Inc. - P.O. Box 1360, Vista, California USA 92085-1360
Secure Reference Library
© Copyright BioMedical Life Systems
All Rights Reserved
Designed by Rudy Interactive